FiberLase™ GYN LAP CO2 Handpieces
Prescriptive information
Refer to the device user manual for complete instructions on device use.
Intended use/indications for use
The FiberLase GYN LAP Handpieces are designed for use with the FiberLase CO2 family of fibers. The fiber and handpieces are intended for use in surgical procedures requiring ablation, vaporization, excision, incision and coagulation of soft tissue. The FiberLase GYN LAP Handpieces are indicated for use in Gynecological procedures such as Endometriosis and Adhesiolysis.
Contraindications
See the user manual for your laser system for a list of contraindications for Gynecological procedures.
Warnings and precautions
- Do not apply laser energy if the fiber’s tip does not protrude from the handpiece tip to avoid damage to the fiber and the handpiece.
- Do not apply laser energy unless the fiber’s tip can be seen protruding from the handpiece tip about 2–3mm.
- Only utilize the laser if the handpiece tip protrudes from the distal end of the suction irrigator.
- When utilizing powers above 25W, lasing for more than 60 seconds continuously may cause excessive heating of the GYN LAP-S handpiece. Be sure to allow sufficient cooling time for this handpiece between contiguous extended lasing sessions.
- Do not active fluid suction or aggressively manipulate tissue while handpiece is extended.
- Be sure to disassemble the handpieces prior to cleaning and sterilization.
- These instruments should be cleaned manually and thoroughly after each use. Automated cleaning alone may not be sufficiently effective.
- Alkaline cleaning agents and those with chlorine or chloride as the active ingredient are corrosive to metal and should not be used. End-of-life for reusable instruments is usually necessitated by normal wear resulting from repeated use.
- Always observe universal precautions when handling contaminated/bio-hazardous materials.
Adverse events/complications
Refer to the Laser Console User Manual for adverse events/complications specific to Gynecological procedures.
Precautions can be found in the product labeling supplied with each device.
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician.
Boston Scientific acquired the global surgical business of Lumenis Ltd.
All trademarks are the property of their respective owners.
URO-1225206-AA JAN 2022
FiberLase Robotic Drop-In Guide — IFU 52064058
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.
Prior to use, please refer to all applicable “Instructions for Use” for more information on Intended Use/Indications for Use, Contraindications, Warnings, Precautions, Potential Adverse Events, and Operator’s Instructions.
Intended use/indications for use
See the user manual for your laser console and delivery system for the intended use/indications for use by specialty.
Contraindications
See the user manual for your laser console and delivery system for a list of contraindications by specialty.
Warnings
- The FiberLase CO2 fiber has bend radius and power limitations that must be respected. To avoid damage to the laser fiber when using the Drop-In Guide, read and thoroughly understand the information in the Instructions for Use for the CO2 fiber, pertaining to minimum bend radius, maximum power and external pressurized purge air use. Be especially mindful of the bends and tension where the Drop-In Guide enters and exits the portal.
Precautions
There is no section or statement in the IFU referring to precautions. See the user manual for your laser console and delivery system for information.
Potential adverse events
Refer to the Laser Console and Delivery System User Manual/IFU for adverse events/complications that may be specific to each surgical specialty.