PERCIVA™ MINI Implantable Cardiac Defibrillator (ICD)
Reimbursement
Configure or select a product to continue to order
- Overview
- Technical specifications
Overview
HeartLogic heart failure diagnostic
HeartLogic uses multiple sensors to track physiological trends, combines them into one composite index and sends a proactive alert of a potential heart failure event weeks in advance.1
ImageReady MR-Conditional Systems
Your patients can safely undergo Full Body MRI scans with ImageReady MR-Conditional Systems.*
Longevity Information
The following tables represent example pulse generator life expectancy estimation (implant to explant) as provided in product labeling. For specific programmable parameter ranges, refer to product labeling at www.bostonscientific.com/manuals, or contact Boston Scientific technical services or your local representative.
| Projected Longevitya | Type | Pacing Amplitude | Pacing | 500Ω with LATITUDEb | 700Ω with LATITUDEb | 900Ω with LATITUDEb | 700Ω no LATITUDE™, RS, or HFSSc |
|---|---|---|---|---|---|---|---|
| VR | |||||||
| Typical programmed setting | VR | 2.5 V | 15% | 8.0 | 8.0 | 8.1 | 9.1 |
| Maximum labeled longevity | VR | 2.0 V / Off | 0% | 8.2 | 8.2 | 8.2 | 9.3 |
| DR | |||||||
| Typical programmed setting | DR | 2.5 V | 15% | 7.2 | 7.3 | 7.3 | 8.1 |
| Maximum labeled longevity | DR | 2.0 V | 0% | 7.5 | 7.5 | 7.6 | 8.5 |
- Assumes 70 PPM LRL; DDDR mode; 0.4 ms Pulse Width (RA, RV); sensors On, Heart Failure Sensor Suite On.
- Projected longevity is calculated assuming 2 maximum energy charging cycles per year, including automatic capacitor re-forms and therapeutic shocks. These calculations also assume 3-channel EGM Onset is on and that the pulse generator spends 3 months in Storage mode during shipping and storage.
a. Assumes ZIP telemetry use for 2 hours at implant and for 40 minutes annually for in-clinic follow-up checks.
b. Assumes standard use of the LATITUDE™ Communicator as follows: Daily Device Check on, quarterly scheduled remote follow ups, and other typical interrogations.
c. Assumes LATITUDE™ Communicator is not used, Respiratory Sensor is Off, and Heart Failure Sensor Suite is Off.
Additional Longevity Information
- Boston Scientific devices have corporate warranties at 5 years in available geographies. Warranty information available at www.bostonscientific.com/warranty.
- Devices use Li/MnO2 chemistry.
- The Usable Battery Capacity is 1.0 Amp-hours (typical implant to battery capacity depleted).
- Shelf life is 2 years (before use by).
Technical specifications
| Model | Type | Size (cm) (W x H x D) | Mass (g) | Volume (cc) | Connector Type | MR Conditional | C-Codes |
|---|---|---|---|---|---|---|---|
| D400 | VR | 5.23 x 7.14 x 0.99 | 61.9 | 28.5 | RV: IS-1/DF-1 | No | C1722 |
| D401 | DR | 5.23 x 7.14 x 0.99 | 62.3 | 28.5 | RA: IS-1; RV: IS-1/DF–1 | No | C1721 |
| D412 | VR | 5.23 x 6.71 x 0.99 | 60.0 | 26.5 | RV: DF4 | Yes | C1722 |
| D413 | DR | 5.23 x 7.03 x 0.99 | 62.5 | 28.0 | RA: IS-1; RV: DF4 | Yes | C1721 |
References
*When conditions of use are met.
1. Boehmer JP, Hariharan R, Devecchi FG, et al. A Multisensor algorithm predicts heart failure events in patients with implanted devices: resultsfrom the MultiSENSE study. JACC Heart Fail. 2017 Mar;5(3):216-25.