Skip to main content

SupraCross™ RF Transseptal Solution

Indications, safety, and warnings

CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician. Rx only. Prior to use, please see the complete “Instructions for Use” for more information on Indications, Contraindications, Warnings, Precautions, Adverse Events, and Operator’s Instructions. 


SupraCross™ RF Wire

INDICATIONS FOR USE

The SupraCross™ RF Wire is indicated for creation of an atrial septal defect in the heart.
 

CONTRAINDICATIONS

The SupraCross™ RF Wire is not recommended for use with any conditions that do not require the creation of an atrial septal defect. The Connector Cable is not recommended for use with any other RF Generator or any other device.
 

WARNINGS

  • Only physicians with a thorough understanding of angiography and percutaneous interventional procedures should use this device.  It is recommended that physicians avail themselves of pre-clinical training, a review of pertinent literature and other appropriate education before attempting new interventional procedures.
  • The SupraCross™ RF Wire and Connector Cable are supplied STERILE using an ethylene oxide process.  Do not use if the package is damaged.
  • Laboratory staff and patients can undergo significant x-ray exposure during RF puncture procedures due to the continuous usage of fluoroscopic imaging. This exposure can result in acute radiation injury as well as increased risk for somatic and genetic effects. Therefore, adequate measures must be taken to minimize this exposure.
  • The SupraCross™ RF Wire and Connector Cable are intended for single patient use only. Do not attempt to sterilize and reuse either devices. Reuse can cause patient injury and/or the communication of infectious disease(s) from one patient to another. Reuse may result in patient complications.
  • The SupraCross™ RF Wire must be used with the Connector Cable provided.  Attempts to use it with other connector cables can result in electrocution of the patient and/or operator.
  • Do not use the SupraCross™ RF Wire with electrocautery or electrosurgery generators, connector cables or accessories as attempted use can result in patient and/or operator injury.
  • The Connector Cable must only be used with the RFP-100A Baylis RF Generator and the included SupraCross™ RF Wire. Attempts to use it with other RF Generators and devices can result in electrocution of the patient and/or operator.
  • The SupraCross™ RF Wire must be used with 0.035” compatible transseptal sheath and/or dilator devices. Use of incompatible accessory devices may damage the integrity of the SupraCross™ RF Wire or accessory devices and may cause patient injury.
  • The SupraCross™ RF Wire has only been validated for transseptal puncture use through SupraCross™ dilators which have been demonstrated to provide the required support for optimal function
  • The active tip and distal curve of the SupraCross™ RF Wire are fragile. Be careful not to damage the tip or the distal curve while handling the SupraCross™ RF Wire.  If the tip or the distal curve becomes damaged at any time during its use, discard the SupraCross™ RF Wire immediately. Do not attempt to straighten the active tip if bent. Damage to device can lead to patient injury.
  • The SupraCross™ RF Wire is not intended for use with neonatal patients (i.e. less than one month of age).  Do not attempt to treat neonatal patients with the SupraCross™ RF Wire.
  • Do not attempt to insert or retract the SupraCross™ RF wire through a metal cannula or a percutaneous needle, which may damage the device and may cause patient injury.

 

PRECAUTIONS

  • Do not attempt to use the SupraCross™ RF Wire and the Connector Cable before thoroughly reading the accompanying Instructions for Use.
  • RF puncture procedures should be performed only by physicians thoroughly trained in the techniques of RF-powered puncture in a fully equipped catheterization laboratory.
  • The sterile packaging should be visually inspected prior to use. Do not use the devices if the packaging has been damaged or compromised.
  • Visually inspect the SupraCross™ RF Wire and Connector Cable prior to use to ensure there are no cracks or damage to the insulating material.  Do not use the wire or the cable if there is any damage.
  • Do not use the SupraCross™ RF Wire and/or Connector Cable after the use-by date indicated on the label.
  • The SupraCross™ RF Wire and Connector Cable are intended for use with only those devices listed in Section VIII, Equipment Required.
  • Read and follow the manufacturer’s Instructions For Use for the DIP electrode.  Always use DIP electrodes that meet or exceed IEC 60601-2-2 requirements.
  • Placement of the DIP electrode on the thigh could be associated with higher impedance.
  • In order to prevent the risk of ignition, ensure that flammable materials are not present in the room during RF power application.
  • Take precautions to limit the effects that the electromagnetic interference (EMI) produced by the Baylis RF Generator may have on the performance of other equipment. Check the compatibility and safety of combinations of other physiological monitoring and electrical apparatus to be used on the patient in addition to the Baylis RF Generator.
  • Adequate filtering must be used to allow continuous monitoring of the surface electrocardiogram (ECG) during RF power applications.
  • Do not attempt to insert and use the proximal end of the SupraCross™ RF Wire as the active tip.
  • Do not bend the SupraCross™ RF Wire or the Connector Cable. Excessive bending or kinking of the wire shaft, distal curve of the wire and/ or the Connector Cable may damage the integrity of the device components and may cause patient injury. Care must be taken when handling the SupraCross™ RF Wire and Connector Cable.
  • Careful manipulation of the SupraCross™ RF Wire must be performed to avoid vessel trauma. If resistance is encountered, DO NOT use excessive force to advance or withdraw the SupraCross™ RF Wire or ancillary sheath and/or dilator assembly. Excessive force may lead to bending or kinking of the device limiting advancement and retraction of sheath and/or dilator device.
  • SupraCross™ RF Wire and ancillary sheath and/or dilator assembly advancement should be done under imaging guidance. The use of visible markers on the wire body are only an approximate guide for positioning the wire tip with the distal end of the dilator.
  • Do not attempt to deliver RF energy until the active tip of the SupraCross™ RF Wire is confirmed to be in good contact with the target tissue.
  • Avoid RF energy delivery of the SupraCross™ RF Wire with incompatible dilator or cannula devices, which may lead to patient burns, ineffective puncture or failure to puncture.
  • It is recommended not to exceed five (5) RF power applications per SupraCross™ RF Wire.
  • Never disconnect the Connector Cable from the Baylis RF Generator while RF power is being delivered.
  • Never disconnect the Connector Cable from the Baylis RF Generator by pulling on the cable. Failure to disconnect the cable properly may result in damage to the cable.
  • Do not twist the Connector Cable while inserting or removing it from the Isolated Patient Connector on the Baylis RF Generator.  Twisting the cable may result in damage to the pin connectors.
  • The Baylis RF Generator is capable of delivering significant electrical power.  Patient or operator injury can result from improper handling of the SupraCross™ RF Wire and/or DIP electrode, particularly when operating the device.
  • During power delivery, the patient should not be allowed to come in contact with ground metal surfaces.
  • Apparent low power output or failure of the equipment to function properly at normal settings may indicate faulty application of the DIP electrode, failure to an electrical lead, or poor tissue contact at the active tip.  Check for obvious equipment defects or misapplication. Attempt to better position the active tip of the SupraCross™ RF Wire against the atrial septum. Only increase the power if low power output persists.
  • If using electroanatomical mapping guidance, it is recommended to use it along with alternative imaging modality in the event there is loss of visibility of the device.

 

ADVERSE EVENTS

Adverse events that may occur while creating an atrial septal defect include:

  • Tamponade
  • Sepsis/Infection
  • Thromboembolic episodes
  • Vessel perforation
  • Atrial Fibrillation
  • Myocardial Infarction
  • Vessel spasm
  • Sustained arrhythmias
  • Atrial Flutter
  • Hemorrhage
  • Vascular thrombosis
  • Perforation of the myocardium
  • Hematoma
  • Allergic reaction to contrast medium
  • Ventricular Tachycardia
  • Pain and Tenderness
  • Arteriovenous fistula
  • Pericardial effusion
  • Tachycardia
  • Vascular Trauma
  • Additional Surgical Procedure
  • Wire entrapment/entanglement
  • Foreign body/wire fracture

 

97186433 (Rev. A)


SupraCross™ Steerable Sheath

INDICATIONS FOR USE

The SupraCross™ Steerable Sheath kit is indicated for introducing various cardiovascular catheters to the heart, including the left side of the heart through the interatrial septum.
 

CONTRAINDICATIONS

There are no known contraindications for this device.
 

WARNINGS

  • Laboratory staff and patients can undergo significant x-ray exposure during interventional procedures due to the continuous usage of fluoroscopic imaging.  This exposure can result in acute radiation injury as well as increased risk for somatic and genetic effects.  Therefore, adequate measures must be taken to minimize this exposure.
  • The SupraCross Steerable Sheath kitis intended for single patient use only. Do not attempt to sterilize and reuse the SupraCross   Steerable Guiding Sheath kit.  Reuse can cause the patient injury and/or the communication of infectious disease(s) from one patient to another.
  • Care should be taken to ensure that all air is removed from the sheath before infusing through the side port.
  • Care should be taken when inserting or removing the dilator and catheters from the sheath.
  • Do not attempt direct percutaneous insertion of the sheath without the dilator as this may cause vessel injury.
  • Damage to guidewire may result if withdrawn through a metal needle cannula.
  • Maintain continuous hemodynamic monitoring throughout procedure
  • Provide continuous heparinized saline infusion while the introducer remains in vessel.
  • To minimize vacuum effects during withdrawal, remove components/aspirate slowly. Refrain from aspiration if a wire is directly through the valve.
  • Avoid contact with liquids other than blood, isopropyl alcohol, contrast solution or saline.
  • Prior to steerable sheath’s delivery and removal, ensure distal section is as straight as possible.
  • Do not kink, stretch or severely bend steerable sheath.
  • Do not use surgical instruments to handle sheath.
  • The sheath device shaft in its entirety is coated with a hydrophobic lubricious coating for smoother device manipulation. The following warning must be considered:
  • Excessive wiping and/or wiping with a dry gauze may damage the coating.
  • The guidewire is coated with a lubricious coating. The following warnings must be considered:
    • Use with incompatible introducers or dilators may affect the device performance and integrity, including coating integrity.
    • Excessive manual bending and/or shaping of the device may affect the coating integrity.
  • DO NOT attempt to insert or retract the guidewire through a metal cannula or a percutaneous needle, which may damage the guidewire and may cause patient injury.

 

PRECAUTIONS

  • Do not attempt to use the SupraCross Steerable Sheath kit before thoroughly reading the accompanying Instructions for Use.
  • Careful manipulation must be performed to avoid cardiac damage, or tamponade. Sheath, dilator and guidewire advancement should be done under fluoroscopic guidance.  If resistance is encountered, DO NOT use excessive force to advance or withdraw the device.
  • The SupraCross Steerable Sheath kit is supplied STERILE using an ethylene oxide process.
  • The sterile packaging and all components should be visually inspected prior to use.  Do not use if the device, packaging or sterile barrier have been compromised or damaged.
  • Only physicians thoroughly trained in the techniques of the approach to be used should perform interventional procedures.
  • Only physicians or personnel trained in aseptic techniques should perform aseptic presentation.
  • Do not use device after its “Use By” date.
  • Avoid deflecting distal end of sheath during delivery and removal, otherwise damage to vessels may occur.
  • The SupraCross Steerable Sheath kit is not compatible with transseptal needles such as the “NRG™  Transseptal Needle”.
  • Do not reshape the distal tip or curve of the guidewire. Excessive bending or kinking of the distal curve may damage the integrity of the wire or coating and lead to patient injury.
  • Only use compatible tip straighteners with the guidewire.
  • Do not attempt to insert the proximal end of the guidewire as the distal end.
  • Confirm ancillary devices are compatible with the dilator and guidewire diameters before use.
  • Individual patient anatomy and physician technique may require procedural variations.
  • Do not attempt to use the guidewire with electrocautery tools.
  • Avoid guidewire contact with liquids other than blood, isopropyl alcohol, contrast solution, or saline.

 

ADVERSE EVENTS

Adverse events that may occur while using the SupraCross™ Steerable Sheath include:

  • Infection
  • Air embolus
  • Local nerve damage
  • Vasovagal reaction
  • Dissection
  • Vessel spasm
  • AV fistula formation
  • Atrial septal defect
  • Pseudoaneurysm
  • Aortic puncture
  • Arrhythmias
  • Perforation and/or tamponade
  • Hematoma
  • Hemorrhage
  • Catheter entrapment
  • Embolic events
  • Stroke
  • Valve damage
  • Myocardial infarction
  • Pacemaker/defibrillator lead displacement
  • Pulmonary edema
  • Coronary artery spasm and/or damage
  • Vessel trauma
  • Pericardial/pleural effusion

 

97186434 (Rev. A)