INGEVITYTM+

Pacing Lead

Closeup of ingevity plus pacing lead for conduction system pacing.
INGEVITY+, active fixation pacing lead

Proven Longevity.
Predictable Performance.

INGEVITY+ pacing leads are 6F (2.0 mm) steroid-eluting, endocardial/interventricular septal pace/sense leads intended for chronic right atrium and/or right ventricle 
or left bundle branch area application.

Product details

 
The INGEVITY+ pacing lead demonstrated 95.6% implant success with left bundle or left septal capture.1

Why choose INGEVITY+


Proven lead performance in a new application.

Low and stable pacing threshold with INGEVITY+ for left bundle branch area (LBBA) implants similar to right ventricle apical locations.
Average pacing capture threshold of 0.94V for INGEVITY+ at 3 month
Mean pacing capture thresholds were 0.94V @ 0.4 ms at 3 months for LBBA2
98.2% of LBBA pacing thresholds were less than or equal to 2V with INGEVITY+
98.2% of LBBA thresholds were less than or equal to 2V at 0.4ms2
97.6% of LBBA pacing leads showed no threshold rise over 1V at 3 months
97.6% of LBBA leads did not have a threshold rise >1V at 3 months2


Tested and proven lead design, expanded to LBBAP

400 million cycle distal tip fatigue testing for use in LBBAP 
  • The stylet design INGEVITY+ pacing lead allows for precise positioning including continuous pacing and monitoring of electrical measurements during lead body rotations3
  • Outer insulation has more than 20 years of experience with Boston Scientific pacing leads and continues to be durable even when exposed to rough simulated LBBAP use3,4,5
  • Designed with 3 robust layers of insulation between conductors6 
  • Zero conductor fractures demonstrated under simulated use conditions with extensive preconditioning, INGEVITY+ leads in LBBA positions met a minimum of a 10-year survivability similar to right ventricular apical pacing4


Enhanced Design

200% tensile strength increase with enhanced design changes. 
Close-up of INGEVITY+ lead showing tri-filar inner coil design for improved torque transmission

Improved inner coil design

The traditional tri-filar inner coil of INGEVITY+ improves the torque transmission of the inner coil and allows for removal of the ETFE covering. The traditional tri-filar inner coil of INGEVITY+ improves the torque transmission of the inner coil and allows for removal of the ETFE covering.

Experience predictable performance with low and consistent turn counts:

  • In tortuous bends
  • In atrial placements
  • When repositioning the lead
  • Across a wider breadth of physician techniques
Coupler and terminal pin joints in INGEVITY+ lead, showcasing enhanced mechanical strength and durability

Improved joint design

  • Incorporated joint design principles from 
the RELIANCE™ family of leads, which have multi-filar coils and proven performance
  • Reduces potential stress concentrations which could cause conductor fracture if the coil was over-torqued or severely bent in a 
fatigue condition
  • Strengthens the joint construction so INGEVITY+ can withstand extraction forces over 200% more than the previous generation

Featured therapy

"INGEVITY+ pacing lead with exposed helix tip and stylet-based components for precise LBBA placement "

Conduction system pacing

Anchored by the INGEVITY+ pacing lead, Boston Scientific’s Conduction System Pacing portfolio works together to precisely navigate to the septum and efficiently fixate with the stylet design allowing continuous monitoring of electrical measurements to confidently capture the left bundle branch area.

Explore therapy details


Products in this therapy

Closeup of ingevity plus pacing lead for conduction system pacing.
INGEVITY+ Pacing Lead
Helix locking tool for conduction system pacing
Helix Locking Tool

Technical specifications

INGEVITY+ leads utilize an IS-1 bipolar connector and are compatible with IS-1 Fixation and helix locking tool1. The tip incorporates an IROX™ (iridium oxide) coating on the tip electrode2. The electrically active helix allows for continuous pacing on the stylet with lead body rotations3. INGEVITY+ can be safely delivered by Boston Scientific’s portfolio of Site Selective Pacing Catheters.4​

INGEVITY+ Pacing Lead Table

Product INGEVITY+ Pacing Lead
Model/Length 7840 / 45 cm
7841 / 52 cm
7842 / 59 cm
Type Bipolar Atrial / Ventricular Straight
Connector IS-1 BI
Compatibility Pulse generators with an IS-1 port, which accepts an IS-1 terminal
MRI Conditions of Use* ImageReady™ MR-Conditional System when used with an MR-Conditional pulse generator
- Full body scan 1.5T and 3T
Introducer without guide wire 6F (2.0 mm)
Introducer with guide wire 9F (3.0 mm)
Fixation Extendable / retractable helix
Expected number of rotations to fully extend/retract the helix** 6 ± 2 turns with straight stylet
7 ± 3 turns with J stylet
Recommended maximum number of turns to extend / retract the helix** 30
Nominal fixation helix penetration depth 1.8 mm
Nominal Electrode:
Fixation helix surface area
Distance between electrodes
Anode electrode surface area
4.5 mm2
10.7 mm
20 mm2
Nominal Diameter:
Insertion
Anode electrode
Lead body
Fixation helix
2.0 mm (6 F)
2.0 mm
1.9 mm
1.2 mm
Material:
External insulation
Internal insulation
Terminal ring contact
IS-1 terminal pin contact
Tip electrode
Anode electrode
Polyurethane (55 D) Silicone rubber
316 L stainless steel 316L stainless steel
IROX™ (iridium oxide) coated Pt-Ir IROX (iridium oxide) coated Pt-Ir
Conductor Type Tri-filar inner coil of MP35N™ and single-filar outer coil of MP35N with a silver core.1 (MP35N is a trademark of SPS Technologies, Inc.)
Steroid 0.91 mg dexamethasone acetate
Radiopaque Markers Pt-Ir
Suture Sleeve Radiopaque white silicone rubber
C-code 1898