DYNAGEN™ X4

Cardiac Resynchronization Therapy Defibrillator (CRT-D)

EasyView™ header with color coded lead ports

Only the X4 quadripolar CRT-D provides 17 pacing vectors plus industry-leading projected longevity for more streamlined patient care.

Key Resources

Explore

 

Product Details

Models G156 and G158

 • The DYNAGEN X4 CRT-D features 17 pacing vector options designed to avoid phrenic nerve stimulation and high thresholds.

• Brings forward the same long lasting battery technology as previous generations of Boston Scientific devices.

• Includes the new EasyView™ header with color coded lead ports, designed to improve implant efficiency. 

• The world’s thinnest high-energy CRT-D on the market.

Offers an uncompromised set of features

  • An advanced system solution for patient comorbidities and HF monitoring (HF Perspectiv™ report, LATITUDE™ NXT Remote Patient Management* with weight scale and blood pressure sensors, and Respiratory Rate Trend). 

  • AcuShock™Advanced Technology, multiple programmable options to reduce inappropriate and unnecessary shocks, including a choice of rhythm discriminators, antitachycardia pacing (ATP) therapy in all rates zones, and advanced sensing and filtering. 

  • RhythmID™ with RhythmMatch™ to allow customization of Rhythm ID algorithm to reduce inappropriate shocks. 

  • Wireless ECG to save time and simplify follow-up. 

  • SafetyCore™ technology intended to provide lifesaving shock therapy and basic pacing functionality in the event of an unrecoverable fault. 
*Caution: Investigational device. Limited by United States Federal Law to investigational use. Not for sale/use in the United States.

Mechanical Specifications

ModelTypeSize (cm)
(W x H x D)
Mass
(g)
Volume
(cm3)
Connector Type
(RA RV LV)
G158X4 CRT-D5.37 x 8.18 x 0.9973.832.5RA:IS-1; RV:DF4; LV:IS4
G156X4 CRT-D5.37 x 8.08 x 0.9973.432.0RA:IS-1; RV:IS-1/DF-1; LV:IS4
Pulse Generator Projected Longevity (All DYNAGEN X4 CRT-D Models a,b,c,d)

 Pacing Amplitude

Projected Longevity (years) at 500Ω and 700Ω
 Pacing Impedance (RV and LV) 
 
RA/RVLV500Ω700Ω
2.5 V3.0 V8.18.6
2.5 V3.5 V7.68.2
3.5 V3.5 V6.87.5
3.5 V5.0 V5.76.5
a Assumes ZIP™ telemetry use for 3 hours at implant time and for 40 minutes annually for in-clinic follow-up checks.
b Assumes standard use of the LATITUDE™ Communicator as follows: Daily Device Check on, monthly Full Interrogations (scheduled remote follow ups, and quarterly patient-interrogations).
c Assumes 60 bpm LRL; DDDR mode; 100% biventricular pacing; 15% atrium pacing and 0.4 ms pacing Pulse Width (RA, RV, LV); RA Impedance 500 Ω; sensors On.
d Projected longevity is calculated assuming 3 maximum energy charging cycles per year, including automatic capacitor re-forms and therapeutic shocks. For the final year of device service, an additional 5 charging cycles are assumed to account for additional automatic capacitor re-forms as the device approaches the Explant indicator. These calculations also assume 3-channel EGM Onset is set to On, and that the pulse generator spends 6 months in Storage mode during shipping and storage.

Longevity Information

• For longevity calculations based on different settings please contact Boston Scientific technical services or your local representative.

• Boston Scientific devices have a limited product warranty of 6 years in available geographies. See www.bostonscientific.com/cardiacrhythm-resources for complete warranty terms and conditions.

• Devices use Li/MnO2 battery chemistry.

• The usable battery capacity is 1.9 Amp-hours (typical implant to battery capacity depleted). 

• Shelf life is 2 years (before use by). 