Supported Devices

MR Conditional Defibrillation Systems and Pacing System Components for 1.5T and 3T

ICDs, CRT-Ds and S-ICDs

Programmer Guide - ICD-CRTD-SICD

Pacing Systems

Programmer Guide - Pacing Systems

Warnings

  • The combined use of a FINELINE II lead and an INGEVITY MRI lead with a Boston Scientific MR Conditional pulse generator has not been evaluated and does not constitute an ImageReady™ MR Conditional Pacing System.

  • Only the combination of INGEVITY MRI lead(s) with an ESSENTIO MRI, PROPONENT MRI, or ACCOLADE MRI pulse generator is valid to use with either 1.5 T or 3T scanners. All other allowable combinations of Boston Scientific MR Conditional system components must use only 1.5 T scanners.
Top