CRT Product Longevity
U.S. Reason for Out of Service

CRM Product Performance Report Q4 2011

COGNIS

Models N118/N119/N120/P106/P107/
              P108

COGNIS
Models N118/N119/N120/P106/P107/
              P108
Worldwide Distribution:86,000
Worldwide Confirmed Malfunctions:124
Without Compromised Therapy With Compromised Therapy Total
Electrical251035
108Safety Core-electrocautery204
110High-voltage capacitor-1
115Low-voltage capacitors5-
119Integrated circuit-5
Mechanical204565
2Subpectoral implant 2009524
101Transformer-5
106Difficulty securing lead98
113Header contacts36
121Header32
Software9-9
114Safety Core-programming1-
117Fault codes not displayed post-EOL2-
120Memory errors6-
Other10515
Non-patterned105
WW Confirmed Malfunctions 6460124

More details about malfunctions

References cited in table above