Boston Scientific Logo

ALTRUA™ Pacemaker Family

The ALTRUA family of pacemakers deliver the benefits of natural pacing by accurately sensing and responding to the body’s signals. AV Search Hysteresis is designed to reduce unnecessary RV pacing without dropping beats, an MV Blended Sensor restores chronotropic competence1, and Stored Onset EGMs collect vital diagnostic data without sacrificing device longevity.

ALTRUA pacemakers provide physiologic pacing that can be tailored to fit the specific needs of patients.
 Contact Us About This Product
HEALTH CARE PROFESSIONALS
Overview

ASSURANCE

Confidence in performance: Peace of mind for you and your patients
  • Engineered through Aurora, Boston Scientific Cardiac Rhythm Management’s continuous-improvement program.
  • Enhanced features and robust design are built on the experience of the INSIGNIA platform, reflecting the best of Boston Scientific’s industry-leading experience.
  • Stored Onset EGMs, nominally ON, means you don’t have to choose between collecting valuable diagnostic information and device longevity.
  • Enhanced header features redesigned set screws, seal plugs, and the addition of MED-420 lubricant designed to improve overall function of the header.

ENHANCED THERAPY

Diagnostic and therapeutic capabilities to tailor therapy to meet specific patient needs

Managing RV pacing
The ALTRUA 60 series offers an AV Delay extendable to 400ms.

  • Multiple programming options provide increased flexibility and are designed to minimize unnecessary RV pacing.
    - Fixed AV Delay: 10ms – 400ms
    - Dynamic AV Delay: Extendable up to 400 ms
    - AV Search Hysteresis (AVSH): Extendable to 400ms maximum, allowing more time for intrinsic ventricular conduction to occur after an atrial event
  • AVSH extendable to 400ms is designed to promote intrinsic conduction without dropping ventricular beats.

Restoring Chronotropic Competence
The MV Blended Sensor provides a physiologic response to movement and breathing, restoring chronotropic competence(1) and providing Life Adaptive Pacing ™

  • MV Blended Sensor can detect changes in motion and minute ventilation (respiratory rate and tidal volume). Accelerometer senses changes in motion but cannot respond appropriately to increased workload in daily activities such as carrying groceries, walking upstairs, and household chores.
    Learn More about MV Blended Sensor and Chronotropic Incompentence »

Collect Valuable Diagnostics
Stored Onset EGMs provides a more complete snapshot of patients’ rhythm before, during, and after an event without sacrificing battery longevity.

  • Using Onset has a negligible impact on device longevity because stored EGMs were integrated into the original device design*.
  • Clinical diagnosis using a stored EGM cannot be achieved in up to 61% of episodes without an onset EGM2.
  • Stored Onset EGMs (nominally ON) make it possible to detect and diagnose events in patients that are asymptomatic.
    *Provides up to 31 events/110 seconds of dual-channel EGM storage with minimal (<0.5) longevity impact.

EXCEPTIONAL EXPERIENCE

Family of devices to meet a variety of patient and physician requirements: For an excellent brady treatment experience, look to the ALTRUA Pacing System from Boston Scientific Cardiac Rhythm Management.

  • The ALTRUA Pacing System portfolio includes ALTRUA 60, 40, and 20 series of devices, offering an array of features and functionality.
  • White seal plugs are designed to provide improved sealing and better visibility for increased efficiency at implant.
  • World-class support from a company you can rely on.



MECHANICAL SPECIFICATIONS

ALTRUA 60 EL

 

 

 

 

 

 

 

Model #

Type

Size (HxWxD)

Mass (g)

Volume (cc)

Connector

Longevity (years)

Warranty (years)

S606

DR

49 x 43 x 8

29.6

12.1

IS-1

8.83

7, 5 full + 2 prorated


ALTRUA 60

 

 

 

 

 

 

 

Model #

Type

Size (HxWxD)

Mass (g)

Volume (cc)

Connector

Longevity (years)

Warranty (years)

S603

DR

44 x 42 x 8

25.4

10.8

IS-1

6.53

5

S602

DR

49 x 43 x 8

29.6

12.6

IS-1 compatible/3.2mm

8.83

5

S601

SR

42 x 42 x 8

23.4

10

IS-1

8.63

5


ALTRUA 40

 

 

 

 

 

 

 

Model #

Type

Size (HxWxD)

Mass (g)

Volume (cc)

Connector

Longevity (years)

Warranty (years)

S404

DR

49 x 43 x 8

29.6

12.1

IS-1

8.14

5

S403

DR

44 x 42 x 8

25.4

10.8

IS-1

5.94

5

S401

SR

42 x 42 x 8

23.4

10

IS-1

7.74

5


ALTRUA 20

 

 

 

 

 

 

 

Model #

Type

Size (HxWxD)

Mass (g)

Volume (cc)

Connector

Longevity (years)

Warranty (years)

S208

DR

49 x 43 x 8

29.6

12.1

IS-1

9.75

5

S205

DR

54 x 49 x 8

32.3

14.9

5/6mm

9.75

5

S203

DR

44 x 42 x 8

25.4

10.8

IS-1

7.15

5

S204

SR

44 x 47 x 8

24.9

11

5/6mm

9.65

5

S201

SR

42 x 42 x 8

23.6

10.1

IS-1 compatible/3.2mm

9.65

5


Boston Scientific Cardiac Rhythm Management: a business reengineered, refocused, and revitalized to serve you better.

C2-194

1 Chronotropic competence is defined by the Model of the Cardiac Chronotropic Response to Exercise, Wilkoff et al. A mathematical model of the cardiac chronotropic response to exercise. Journal of Electrophysiology 1989:3: 176-180. Refer to the Physician's System Guide for more information on adaptive-rate therapy. Additional clinical performance was assessed using INSIGNIA Ultra clinical data with the AutoLifestyle feature programmed On. Data on file.* Boston Scientific Corporation. ALTRUA Pacemaker System Guide. 2008. pg 1-20 – 1-25
2 Nowak B, J. Sperzel, F. Rauscha, L. Mont, J. Alzueta, P. J. Senden, and J. Meyer. Diagnostic value of onset-recordings and marker annotations in dual chamber pacemaker stored electrograms. Europace, 2003; 5: 103-109
3 Longevity projections as described in device user manual. Settings: 60ppm, A = 2.5V, V = 1.0V, 500 ohms, 100% paced, MV Blended Sensor ON, Onset EGMs ON, Automatic Capture ON.
4 Longevity projections as described in device user manual. Settings: 60ppm, A = 2.5V, V = 1.0V, 500 ohms, 100% paced, MV Blended Sensor ON, Onset EGMs ON.
5 Longevity projections as described in device user manual. Settings: 60ppm, A = 2.5V, V = 1.0V, 500 ohms, 100% paced, Accelerometer ON, Onset EGMs ON.


Image Gallery
ALTRUA™ 20, Model S203 ALTRUA™ 40, Model S403 ALTRUA™ 60, Model S603

Product Type

Pacemaker

Used in Procedures

Pacemaker Implantation

Related Medical Areas

Cardiac Rhythm Management

Related Conditions

Heart Conditions